38
B.
Press CONFIG and then select a function as you wish from DCV, DCI (SHIFT +DCV),
Ω2, Ω4 (SHIFT + Ω2), FREQ and PERIOD (SHIFT + FREQ).
Press ENTER on RESOLUTION submenu. If it is not shown on the display, use
◁
and
▷
to locate it.
Use
◁
and
▷
to find your desired resolution. Press ENTER on your choice.
※
Note: The resolution of AC measurements is fixed at 6 1/2 digit.
※
Note: The options for frequency and period measurements are 4 1/2 (slow), 5 1/2
(slow) and 6 1/2 (slow).
Figure 4-3
Remote Interface Operation
Use the following commands on your PC terminal to set the resolution for your
measurement.
CONFigure:<function> <range>,<resolution>
MEASure:<function>? <range>,<resolution>
SENSe:<function>:RESolution <resolution>
Users can specify the resolution in the same units as the measurement function, not in
number of digits. For instance, specify the resolution in volts and the frequency in herts.
The examples are in the following.
CONF:VOLT:DC 10,0.001 (4 1/2 digits on the10Vdc range)
MEAS:CURR:AC? 1,1E-6 (6 1/2 digits on the1A range)
CONF:FREQ 1kHz,0.1Hz (1000 Hz input, 0.1Hz resolution)
VOLT:AC:RES 0.05 (50mV resolution on the ac function)
or
Содержание DL-2060
Страница 1: ...B71 0110 41 INSTRUCTION MANUAL 6 1 2 DIGITAL MULTIMETER DL 2060 DL 2060G DL 2060VG DL 2060R DL 2060VR...
Страница 37: ...31 Figure 3 17 1 2 3 5 5 4...
Страница 135: ...129 C Tree Diagram Please refer to G Notice of Scanner Card if your firmware version is 1 05 or upward...
Страница 136: ...130...
Страница 137: ...131...
Страница 138: ...132 D DL 2060 s Dimension...
Страница 152: ...7F Towa Fudosan Shin Yokohama Bild 2 18 13 Shin Yokohama Kohoku ku Yokohama Kanagawa 222 0033 Japan http www texio co jp...