
30
2.7.2.2 3-Wire RTD Measurements
How to measure temperature with 3-Wire RTD
The following
Figure 3-15
shows theory diagram of 3-Wire RTD measurement.
①
Use terminals switch to select front terminals.
②
Insert a specified adapter into the front terminals.
③
Configure sensor type and transducer as you would be with 4-wire RTD and unit using
TEMP and
◁
or
▷
. When ready, press ENTER button.
④
Press TEMP button.
⑤
Place RTD in the desired position and take the reading on the display.
※
Note: When users do the 3-wire RTD measurement the input LO and sense LO have
to be connected in short condition.
※
Note: “Source” means RTD Probe.
Figure 3-15
2.7.2.3 4-Wire RTD Measurements
How to measure temperature with 4-Wire RTD
The following
Figure 3-17
shows theory diagram of 4-Wire RTD measurement.
①
Use terminals switch to select front terminals.
②
Insert a specified adapter into the front terminals.
③
Configure sensor type,unit
and transducer using TEMP and
◁
or
▷
. When
ready, press ENTER button.
④
Press TEMP button.
⑤
Place RTD in the desired position and take the reading on the display.
1
2
3
5
5
4
In Short Condition
Содержание DL-2060
Страница 1: ...B71 0110 41 INSTRUCTION MANUAL 6 1 2 DIGITAL MULTIMETER DL 2060 DL 2060G DL 2060VG DL 2060R DL 2060VR...
Страница 37: ...31 Figure 3 17 1 2 3 5 5 4...
Страница 135: ...129 C Tree Diagram Please refer to G Notice of Scanner Card if your firmware version is 1 05 or upward...
Страница 136: ...130...
Страница 137: ...131...
Страница 138: ...132 D DL 2060 s Dimension...
Страница 152: ...7F Towa Fudosan Shin Yokohama Bild 2 18 13 Shin Yokohama Kohoku ku Yokohama Kanagawa 222 0033 Japan http www texio co jp...