40
Remote Interface Operation
The automatic input resistance mode can be enabled / disabled. With AUTO OFF
(default), the input resistance is fixed at 10MΩ for all ranges. With AUTO ON, the input
resistance is set to >10GΩ for the three lowest DC voltage ranges (100mV, 1V and 10V).
Use the following command from your PC terminal to disable the auto input DC
resistance setting (the result is a fixed input DC resistance at 10M Ω for all
measurements.)
INPut:IMPedance:AUTO {OFF | ON}
3.1.5 Threshold Resistance (Continuity)
Definition
When testing continuity, the beeper goes off when the measured resistance is less than
the threshold resistance. The threshold resistance can be set to any value between 1Ω
and 1000Ω.
Default
The factory default for continuity threshold resistance is 10Ω. User’s selection is stored
in a volatile memory and the default value will be restored after the meter has been
turned off.
How to set the threshold resistance
User can change the threshold resistance only through the front panel.
Press CONFIG
button and then CONT button. Use
◁
and
▷
buttons to move through the digits and
lower or increase the number to your desired value using
△
and
▽
buttons , then press
ENTER button to set the value. The locations of these buttons are shown with red
rectangle frames in
Figure 4-5
.
Figure 4-5
+
Содержание DL-2060
Страница 1: ...B71 0110 41 INSTRUCTION MANUAL 6 1 2 DIGITAL MULTIMETER DL 2060 DL 2060G DL 2060VG DL 2060R DL 2060VR...
Страница 37: ...31 Figure 3 17 1 2 3 5 5 4...
Страница 135: ...129 C Tree Diagram Please refer to G Notice of Scanner Card if your firmware version is 1 05 or upward...
Страница 136: ...130...
Страница 137: ...131...
Страница 138: ...132 D DL 2060 s Dimension...
Страница 152: ...7F Towa Fudosan Shin Yokohama Bild 2 18 13 Shin Yokohama Kohoku ku Yokohama Kanagawa 222 0033 Japan http www texio co jp...