5 First steps
33
Charging via mains unit (Art.-No. 0554 1096)
✓
The Control Unit is switched off.
1.
Connect the plug of the mains unit to the mains unit socket on
the Control Unit.
2.
Connect the mains plug of the mains unit to a mains socket.
- The charging process starts. The charge condition will be
shown on the display.
- Once the rech. batt. has been charged the instrument will
automatically change to trickle charge.
Charging via meas. box
✓
Control Unit is locked to meas. box or is connected via the
Testo Data bus cable.
✓
The meas. box is supplied via the mains unit.
During operation with low charge power or in switched off state.
5.2.1.2.
Charging the rech. batt. of the meas. box
The rech. batt. pack can only be charged at an ambient
temperature of ±0...+35°C. If the rech. batt. has been discharged
completely, the charging time at room temperature is approx. 6h.
✓
The meas. box is switched off.
> Connect mains cable to meas. box and mains socket.
5.2.1.3.
Battery care
> Do not fully exhaust rechargeable batteries.
> Sore rech. batts. only in charged condition and at low
temperatures, but not below 0°C.
> For longer breaks you should discharge and recharge the
batteries every 3- months. Trickle charging should not exceed 2
days.
5.2.1.4.
Mains operation
In case of danger the instrument must be disconnected from the
electric power supply by simply pulling out the mains cable.
> Always position the instrument so that the power supply plug
can be easily reached.
Control Unit
1.
Connect the plug of the mains unit to the mains unit socket on
the Control Unit.
2.
Connect the mains plug of the mains unit to a mains socket.