7 Maintaining the product
86
4.
Fit a new probe module on the handle and engage in place.
7.9.
Condensate trap / condensate container
With the gas preparation option fitted, the condensate is separated
from the measuring gas and is led into a condensate container that
is isolated from the gas path. In the case of longer measurements
with moist flue gas, the condensate can be led off using a tube
without any external air being carried along.
The filling level of the condensate trap can be read from the
markings.
Emptying the condensate trap / condensate container
The condensate consists of a weak mix of acids. Avoid skin
contact. Make sure that the condensate does not run over
the housing.
CAUTION
Condensate entering the gas path.
Damage to sensors and flue gas pump
!
> Do not empty the condensate trap / condensate container while
the flue gas pump is in operation.
1.
Unlock the condensate trap / condensate container by the
orange handle on the underside.