7 Maintaining the product
94
4.
Replace the exhausted filter with a new filtration non-woven.
5. Close the cover.
6.
Plug on the hose.
7.
Plug the condensate container on the measuring box.
7.13.
Recommended maintenance cycles
Component
Service life
Remedy
Main gas pump
2500h
Renew the pump
Special main gas
pump for long-term
measurement
10000h
Renew the pump
Rinsing/delivery pump 2500h
Renew the pump
Condensate pump
(gas cooler option)
2500h
5000h
Renew pump head
with hose
Renew the pump
Fleece in gas cooler
(gas cooler option)
1200h
Clean housing,
renew fleece