6 Using the product
63
>
[ ]
→
Sensor settings
→
[OK]
→
Calibration /Adjust
→
[OK]
.
Possibly:
> Enter the password:
[Enter]
→ Enter password →
[Next]
→
[OK]
.
- Gas zeroing (30s).
Perform calibration / adjustment of CO-, SO
2
-, NO
2
-, NO-, O
2
-,
H
2
S-, CxHy-sensors:
WARNING
Dangerous gases
Danger of poisoning!
> Observe safety regulations / accident prevention regulations
when handling test gas.
> Use test gases in well ventilated rooms only.
Application of test gas via service adapter (0554 1205) is
recommended, or apply test gas directly to the probe tip to
avoid possible absorptions in the gas path.
1.
Select the parameter:
[▲]
,
[▼]
→
[OK]
2.
[Change]
→ Enter the
test gas concentration (nominal value).
3.
Apply test gas to the sensor.
4.
Start calibration:
[Start]
5. Once the actual value is stable, apply the nominal value (i.e.
adjust the sensor to the nominal value) via
[Adjust]
. The values
are stored in the record.
- or -
Carry out a comparison (calibration) between nominal and
actual value via
[Calibr.]
without adjusting the sensor, and the
values are stored in the record.
- or -
Cancel (no adjustment/calibration):
[esc]
6. Save changes:
[Finished]
Perform calibration / adjustment of the CO
2
-(IR) sensor
Check the CO
2
-(IR)-sensor with the absorption filter to obtain
accurate readings. The displayed CO
2
-value should be
<0.03%CO
2
. If the value is higher, perform calibration and gradient
adjustment.