6 Using the product
66
Switching negative values on/off
1.
[Change]
2.
Select setting:
[▲]
,
[▼]
3.
Confirm the entry:
[OK]
6.1.5.
Programs
Five flue gas measuring programs can be set, saved and executed.
The
Trigger
function (trigger signal as start/stop criterion) is only
available for devices with the trigger input option.
Instrument settings cannot be changed if a program is
active or running.
The program
Flue Gas ( after cat)
checks
whether the meas. box is equipped with a fresh air valve. If
not, a measuring program with normal flue gas
measurement will be added, instead of the program
Flue
Gas ( after cat)
. A program
Flue Gas (
after cat)
without fresh air valve does not show any
sensible measuring results.
Calling up the function:
>
[ ]
→
Programs
→
[OK]
.
Activating / deactivating a program:
>
Select the program:
[▲]
,
[▼]
→
[Enable]
or
[Disable]
.
- When activating a program: The program is activated and the
measurement type matching the program is opened.
Editing the measuring program:
Adjustable parameters:
Parameter
Function
Measurement
program
Edit program name
Measurement
type
Select flue gas menu:
• Flue gas
• Flue gas + m/s
• Flue gas
Δ
P
• Flue gas (before and after catalyst)
• Solid fuel