6 Using the product
51
Factor
Ratio of diluting gas: Measuring gas
x 1
no dilution
x 2
1 : 1
x 5
4 : 1
x 10
9 : 1
x 20
19 : 1
x 40
39 : 1
Auto dilution
4 : 1
If the dilution stage
auto-dilution
is selected, dilution (5x) is
activated automatically when the set switch-off threshold of the
sensor in slot 6 is reached.
• If the surrounding air contains interfering gases, push
the hose onto the dilution inlet and place in a clean
atmosphere.
• If gas from a gas cylinder is used, observe a max.
pressure of 30hPa
• Diluting also changes the resolution of the reading
display, e.g.:
Undiluted resolution 1ppm, with factor 10 resolution
10ppm.
Calling up the function:
>
[ ]
→
Device settings
→
[OK]
→
Dilution
→
[OK]
1.
Single slot
→
[Change]
2.
Set the dilution factor:
[ ]
,
[ ]
.
3.
Confirm the entry:
[OK]
.
Option:
> Without dilution: Press
[Without]
.
Measurement range extension with fixed dilution factor (x5) for
all sensors (fresh air valve option)
The selection of
to dilute all
(x5) results in the dilution of all
sensors (x5). The measuring channels O
2
, CO
2
-(IR), CO
2
, qA,
Lambda, Eta and all measuring channels for flow measurements
are faded out in case of dilution
to dilute all
. 1x deactivates the
dilution (extension of measuring range).
It is possible to calibrate / adjust with test gas when dilution is
switched on to eliminate any measuring errors caused by dilution
(see Calibration / adjustment page 61).
Calling up the function: