5 First steps
43
When an IRDA printer 0554 0549 is used, the control unit
must not be placed on the analyzer box during the printout.
Only readings, which have a display field in the
measurement view assigned, will be saved / printed out.
The measurement data can be printed out parallel to the
saving process, while a measurement program is running.
Measure values of diluted sensors (with enabled
measurement range extension) underlined on the printout.
5.2.9.
Search for meas. boxes
(only available via
Control Unit
tab)
>
[ ]
→
Search for boxes
→
[OK]
.
- Meas. boxes connected via Testo Data bus: are displayed
(tabs)
- Meas. boxes connected via Bluetooth
®
:
• Meas. box found:
Meas. box and Control Unit are connected automatically
• several meas. boxes found:
The available meas. boxes are displayed for selection
An existing Bluetooth
®
connection is disconnected by
selecting a new meas. box from the selection field.
5.2.10.
Confirming an error message
If an error occurs, an error message is shown in the display.
> Confirming an error message:
[OK]
.
Errors which have occurred and have not yet been rectified are
indicated by a warning symbol in the status bar.
Not yet rectified error messages can be displayed in the menu
Error diagnosis
, see Sensor diagnosis, page 48.
5.2.11.
Switching off
Unsaved readings will be lost when the flue gas analyser is
switched off.
Rinse phase
When switched off, the meas. box checks whether flue gases are
still in the sensors. The sensors are rinsed with fresh air, if this
should be necessary. The duration of the rinse phase depends on
the gas concentration in the sensors.