3 Specifications
16
Measurement
parameter
Accuracy
Response
time
CO, H
2
-comp.
±10ppm (0…199ppm)
±5% of reading (200…2000ppm)
±10% of reading (rest of range)
< 40s (t90)
CO
low
, H
2
-comp. ±2ppm (0…39.9ppm CO)
±5% of reading (rest of range)
< 40s (t90)
NO
±5ppm (0…99ppm)
±5% of reading
(100…1999.9ppm)
±10% of reading (rest of range)
< 30s (t90)
NO
low
±2ppm (0…39.9ppm)
±5% of reading (rest of range)
< 30s (t90)
NO
2
±5ppm (0…99.9ppm)
±5% of reading (rest of range)
< 40s (t90)
SO
2
±5ppm (0…99ppm)
±5% of reading (100…1999ppm)
±10% of reading (rest of range)
< 30s (t90)
H
2
S
±2ppm (0…39.9ppm)
±5% of reading (rest of range)
< 35s (t90)
CO
2
-(IR)
±0.3Vol.% ±1% of reading
(0…25Vol.%)
±0.5Vol.% ±1.5% of reading
(rest of range)
< 10s (t90)
heat-up time:
< 15min
CxHy
±400ppm (100…4000ppm)
±10% of reading (rest of range)
< 40s (t90)
Differential
pressure 1
±0.03hPa (-2.99…2.99hPa)
±1.5% of reading (rest of range)
-
Differential
pressure 2
±0.5hPa (-49.9…49.9hPa)
±1.5% of reading (rest of range)
-
Absolute
pressure
±10hPa
-
Type K (NiCr-Ni) ±0.4°C (-100 to 200°C)
±1°C (rest of range)
-
Type S
(Pt10Rh-Pt)
±1°C (0 to 1760°C)
-
Содержание 350 Maritime V2