7 Maintaining the product
89
3.
Open the cover of the service compartment (locking clip) on the
back of the meas. box.
1 Condensate pump
2 Main gas pump
3
Rinsing / feed pump for diluting gas
7.11.1.
Cleaning the main gas pump
1.
Pull the gas pump upwards out of the gas measuring block.
2.
Pull the inlet and outlet hoses off the sockets on the pump head
3.
Loosen the plug connectors and remove the main gas pump.
4.
Loosen the 4 fastening screws (Torx spanner T 9) on the pump
head of the main gas pump.
5.
Pull off the pump head.
6.
Remove the two circlips from the depressions of the pump head
(front and rear).
7.
Remove and clan the pump diaphragm (e.g. white spirit).