6 Using the product
50
6
Using the product
6.1.
Performing settings
6.1.1.
Assigning the right hand function key
The right function key can have a function from the
Options
menu
assigned to it. The menu
Options
is accessed via the left function
key and is available in many different menus. This assignment is
only valid for the currently opened menu / the opened function.
✓
A menu / function is opened in which the
Options
menu is
displayed on the left function key.
1.
Press
[Options]
.
2.
Select option:
[ ]
,
[ ]
.
Depending on the menu / function from which the
Options
menu
was opened, various functions are available.
3.
Assign the selected function to the right function key: Press the
[Config. Key]
.
6.1.2.
Instrument settings
6.1.2.1.
Dilution
(only available via tab
Meas. Box
and with the measurement range
extension option)
Option dilution (for single slot with selectable dilution factors
With active dilution the measuring gas for the sensor in slot 6 is
diluted with ambient air (other possibility: nitrogen gas) in a
controlled manner. For this purpose, the diluting gas is drawn
through a separate gas inlet by a pump and a valve operating on
the principle of pulse width modulation. A filter is installed to protect
the gas path against dust.
If the measurement range extension is active this is indicated by a
clearly noticeable clicking of the valve. In addition the symbol 1/x
appears at the right hand top of the display (in the header) and the
selected dilution factor appears at the corresponding parameter
(the complete line of the diluted parameter appears against a blue
background).
The following dilution factors can be manually set: