6 Using the product
67
Parameter
Function
Reading per
mean value
Mean value
Yes
:
the measurement system saves one reading per
second (measurement rate cannot be changed).
At the end of the gas phase, a mean value is
calculated from the individual values saved thus
far and stored.
Mean value
No
: when the end of the gas phase
is reached, the values measured at this time are
stored.
Start
Determine the start criterion
•
The measuring program is started at any
time (the function key automatically changes
to the stop function).
• Time
Start of measurement at a pre-programmed
time.
• External signal
Trigger signal to control the start of
measuring programs.
• Multi-point
You can manually save multiple readings in a
report while a measurement program is
running, for example from different
measuring points or from different load
levels. In so doing, it is not possible to
change the gas time, mean value, rinsing
time and measurement rate. The load level
or the firing rate for the individual measuring
points can for instance be pre-selected in the
display under
Options
.