![Terumo BCT Trima Accel Скачать руководство пользователя страница 82](http://html1.mh-extra.com/html/terumo-bct/trima-accel/trima-accel_service-manual_1090436082.webp)
Table 2-6: Centrifuge door and basin components (continued)
Name
Function
6
Centrifuge arm
The centrifuge arm holds the rotating centrifuge tubing in place while
the centrifuge spins.
7
Lower bearing holder
The lower bearing holder holds the centrifuge tubing in place at the
lower end of the centrifuge arm.
8
Filler assembly
The filler assembly holds the channel portion of the tubing set in place.
9
Centrifuge loading port
The centrifuge loading port provides an opening in the centrifuge
housing for loading the channel.
Figure 2-18: Centrifuge system block diagram
Starting with an input signal, the control CCA sends a drive command to the centrifuge motor
controller (CMC) after the safety CCA has turned on the 64-volt switch to make centrifuge drive voltage
available to the CMC. The conditions that must be met for the safety CCA to enable the 64-volt switch
include: the door lock solenoid must be unpowered, the door closed sensors must read closed, and the
leak detector must be dry. The control CCA sends a DC voltage (0 V to 5 V) to the CMC as a drive
command. The CMC then converts that voltage into 3-phase drive by superimposing a duty cycle onto
the raw 64-volt-switched voltage on each line, causing the centrifuge motor to spin. Hall-effect sensor
feedback is sent directly to the safety CCA, the control CCA, and the CMC through simple splitting.
The safety and control CCAs have to agree within a certain range of rpm or time, or the centrifuge drive
power is cut. The Hall-effect sensor output is a pulse voltage. If voltage, current, or something in
processing is wrong in the CMC, it sends a fault signal (commutation, direction, or processing) to
Centrifuge System
Trima Accel
®
Automated Blood Collection System • Service Manual
2-21
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...