Mechanical Systems
There are several subsystems that do not have electronic components to them and are purely mechanical.
The IV pole uses a lever system to bind the pole at the desired height position. To move the pole, a
button is pushed that releases this binding, allowing for free movement. A square block inside of the IV
pole tube keeps the IV pole from rotating.
The brake pedal has three positions: 360-degree caster rotation with rolling wheels, no caster rotation
with rolling wheels, and no caster rotation with locked front wheels. The casters are adjustable for
braking purposes.
The display arm has two versions: the newer version uses two adjustable ball detents to keep the display
arm in position, and a previous version uses a push button to adjust the display arm position.
The front door latch plate is adjustable to ensure a seal with the basin front edge.
The body panels are removable to allow access to device systems and parts. There are two side panels, a
front panel, and a back panel.
Operational Description
1-52
Trima Accel
®
Automated Blood Collection System • Service Manual
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...