Valves Test Substate
The RBC, plasma, and platelet valves are moved sequentially to all three of their possible positions. If any
of the positions cannot be reached within 10 seconds, an alarm is generated and the Trima Accel system
resumes at the beginning of the Startup Tests state. If this test passes, the Trima Accel system advances to
the Leak Detector Test substate.
Leak Detector Test Substate
The leak detector output voltage is checked to verify that it is open and can detect fluid leaks. If this test
fails, an alarm is generated. If this test passes, the Trima Accel system advances to the CentShutdown
substate.
CentShutdown Test (Centrifuge Shutdown) Substate
The control computer first checks the 64-volt DC to see if it is within range. Then the Trima Accel
system commands the centrifuge to zero and verifies immobility for two seconds. If this test fails, an
alarm is generated. If this test passes, the Trima Accel system advances to the Door Latch Test substate.
Door Latch Test Substate
These tests check that the door can be seen as both open and closed by the door lock optical and Hall
sensors. The 24-volt switch dedicated to the door solenoid is also checked to ensure that it can be turned
off. If these tests fail an alarm is generated. If these tests pass, the
Load System
button is enabled. There is
also a test to check the ability of the software to lock and unlock the door without power. This test locks
the door and waits to see if it locks even when the power is supposed to be off. If the door is in the locked
state, the alarms are given and the tests are tried again.
GUI Started (Load System Button) Substate
This substate is not a test but instead enables the
Load System
button and waits for the operator to press it
and the
Continue
button from the subsequent screen. The
Donor Info
button also activates at this time.
Load Cassette Substate
After the operator presses the
Load System
button, the cassette is checked for the up position as seen by
the optical sensor. The valves are also commanded to the open position. If they do not achieve the open
position in ten seconds, an alarm is generated.
Table 3-134: Load Cassette Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
0
0
0
0
0
Open
Open
Open
0
Exit Condition
When the operator touches the
Continue
button.
Version 6.1 Software Description
Trima Accel
®
Automated Blood Collection System • Service Manual
3-87
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...