The control CCA sends 5 V to the leak detector CCA and is looking for 2.5 V coming back. The control
CCA reads the difference between the two voltages. A range near 2.5 V is defined as good. Any voltage
below that indicates a leak in the basin. This leak condition causes the control CCA to remove the 64-
volt power to the centrifuge motor, causing the centrifuge to stop.
Sensor System
Trima Accel
®
Automated Blood Collection System • Service Manual
2-17
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...