Table 4-16: Possible door lock electrical failures (continued)
Symptom
Possible Causes
Stuttering or sticking locking shaft
movement, or no locking shaft
response to commands
• Remove the door lock assembly and measure the solenoid coil resistance
while gently manipulating the assembly to try and find an open in the coil.
• Measure the solenoid voltage pulse several times to ensure it is
approximately ±24 V.
• As you measure the solenoid switch voltage, toggle the STS “Door
Solenoid Power” on and off. Ensure that the solenoid switch voltage is
consistent during switching.
• Measure the voltage of the 24 V solenoid switch, the 24 V source, and the
24 V switched on the motor driver CCA to ensure they are all approximately
± 24 V.
• Check the centrifuge motor system for rpm feedback problems. If the
centrifuge is reporting speeds above 60 rpm, the system will not unlock the
solenoid.
Door System Troubleshooting
Trima Accel
®
Automated Blood Collection System • Service Manual
4-21
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...