Positioning the Linear Actuator Sensors
The linear actuator sensors may need to be repositioned if there are cassette alarms or if the machine fails
the cassette plate test in Autotest.
The panel on the same side of the device as the IV pole must be removed.
1.
Boot the device to Service mode and connect with the service software.
2.
Enable the 24 V switch.
3.
Raise the cassette plate.
4.
The two optical sensors are mounted below the pump panel on one of the square members that
house the linear actuator guides. Loosen the lower accessible screw for the top optical sensor.
Figure 5-4: The lower accessible screw (circled)
5.
Place a 0.035 in (0.89 mm) or 0.036 in (0.91 mm) feeler gauge under the upper left corner of the
cassette plate.
6.
Command the cassette plate down and verify that the service software cassette position indicator
status is
Unknown
.
•
If the service software cassette position indicator status is
Down
•
If the service software cassette position indicator status is
Up
•
If the service software cassette position indicator status is
Unknown
Maintenance and Calibration
5-8
Trima Accel
®
Automated Blood Collection System • Service Manual
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...