Blood Run Recovery Substates
The Blood Run state contains eleven recovery substates. Recovery substates interrupt the procedure and
suspend the procedure time clock.
When the recovery routine completes, the Trima Accel system returns to the procedure substate that it
was in prior to recovery. These recovery substates can occur at any time during the Blood Run state.
Only one recovery substate happens at a time. Descriptions of the recovery substates follow.
RBC Spillover (Spillover Recovery) Substate
Spillover recovery can be entered via a command from the operator (by pressing the
Spillover
button on
the Adjust screen) or automatically when detected by the RBC detector. This substate triggers the WBC
Chamber Purge state followed by the Clear Line state (see the description for these substates below). At
the end of spillover recovery, the RBC detector checks for the presence of a spillover an regenerates the
recovery procedure if a spillover persists.
Pumps Pause (Pause Recovery) Substate
This recovery occurs when the pumps are paused for more than 6 seconds. After the inlet pump processes
a predetermined inlet volume, the pumps return to their original collection values and the Trima Accel
system returns to the previous blood run substate.
Air Block (Air Block Recovery) Substate
This recovery is initiated when the operator presses the
Air in Plasma Line
button on the Adjust screen.
The centrifuge speed lowers to a predetermined speed, all valves close, and the pump speeds lower to
remove the air block. The plasma pump runs at a predetermined speed, maintaining chamber flow, until
a predetermined volume of plasma has been pumped. The system then transitions to the Centrifuge
Stop substate.
Centrifuge Stop (Centrifuge Stop Recovery) Substate
This recovery procedure occurs anytime after the centrifuge stops (or drops below a predetermined
speed). This recovery procedure takes precedence over the other recovery procedures. The centrifuge
ramps up to its final commanded speed while the pumps are maintained at low speeds to prevent the
occurrence of a spillover or air block. When the centrifuge reaches the commanded speed plus 60 more
seconds, the pumps return to their normal collection speeds.
Centrifuge Slow (Centrifuge Slow Recovery) Substate
This recovery substate occurs if the centrifuge actual speed drops below a predetermined speed, but is
greater than a predetermined speed. The centrifuge ramps up to 3000 rpm and this recovery continues
for 15 more seconds.
Software Description
3-112
Trima Accel
®
Automated Blood Collection System • Service Manual
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...