Extended Plasma (Extended Plasma Collection) Substate
This substate occurs after all platelets are collected and additional plasma is necessary to achieve the
plasma target. The platelet valve closes and the plasma valve moves to the collect position. After the
plasma target is achieved, the Trima Accel system advances to the RBC Collect substate if RBC is
chosen.
Table 3-38: Extended Plasma Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
Qac
Qin
Qplas
Qplat
Qret
Return
Collect
Return
3000
Exit Condition
When the target (donor-specific) volume is collected in the plasma bag.
Extended Plasma (Extended Plasma Collection) Substate
This substate occurs after all platelets are collected and additional plasma is necessary to achieve the
plasma target. The platelet valve closes and the plasma valve moves to the collect position. After the
plasma target is achieved, the Trima Accel system advances to the RBC Collect substate if RBC is
chosen.
Table 3-39: Extended Plasma Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
Qac
Qin
Qplas
Qplat
Qret
Return
Collect
Return
3000
Exit Condition
When the target (donor-specific) volume is collected in the plasma bag.
RBC Collect (RBC Collection) Substate
If RBC collection is chosen, this substate occurs after all platelets and plasma are collected. The RBC
valve opens until the RBC volume target is achieved. The inlet flow is limited by the configured AC ratio
for RBC procedures.
Table 3-40: RBC Collect Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
Qac
Qin
Qplas
Qplat
Qret
Return
Return
Collect
3000
Exit Condition
When the target (donor-specific) volume is collected in the RBC bag.
Software Description
3-30
Trima Accel
®
Automated Blood Collection System • Service Manual
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...