![Terumo BCT Trima Accel Скачать руководство пользователя страница 240](http://html1.mh-extra.com/html/terumo-bct/trima-accel/trima-accel_service-manual_1090436240.webp)
Pre-Platelet Plasma (Pre-Platelet, Collecting Plasma) Substate
This substate occurs only in procedures where both platelets and plasma are collected. The plasma valve
is commanded to the collect position and the plasma pump speed changes to regulate the RBC line
hematocrit. The AC and inlet pump speeds change to keep the AC infusion rate at the run configured
value. After enough inlet volume has been processed to stabilize the LRS chamber, the Trima Accel
system advances to the Platelet Plasma substate or the Pre-Platelet No Plasma substate.
Table 3-172: Pre-Platelet Plasma Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
Qac
Qin
Qplas
Qplat
Qret
Return
Collect
Return
Qcent
Exit Condition
When the target (donor-specific) volume is processed by the inlet pump.
Pre-Platelet No Plasma (Pre-Platelet, No Plasma Collect) Substate
This substate occurs after the Channel Setup substate, in procedures where platelets, but not plasma, are
collected. The AC and inlet pump speeds change to keep the AC infusion rate at the run configured
value. After a predetermined inlet volume has been processed, the Trima Accel system advances to the
Mid Run substate.
Table 3-173: Pre-Platelet No Plasma Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
Qac
Qin
Qplas
Qplat
Qret
Return
Return
Return
Qcent
Exit Condition
When the target (donor-specific) volume is processed by the inlet pump.
Version 6.1 Software Description
Trima Accel
®
Automated Blood Collection System • Service Manual
3-107
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...