Note:
The door lock solenoid and the two optical sensors in the door lock assembly can all be replaced individually.
Table 4-15: Possible door lock mechanical failures
Symptom
Inspection Points and Possible Solution
Stuttering or sticking locking shaft
movement
• Inspect the door lock assembly and the internal mechanisms for debris,
dried spills, or damaged components. When the locking shaft is moved by
hand, it should move freely once outside of the magnetic force.
• Loosen the four screws holding the solenoid to the assembly, and align the
solenoid to the locking shaft by feel. Replace as necessary.
• Check the locking shaft hole in the latch plate to ensure that the locking
shaft can go through. The hole should be an unbent, clean circle. If
someone attempts to close the centrifuge door when the locking shaft is in
the locked position, the latch plate impacts the locking shaft and can
deform the lock hole. This may also damage the locking shaft. Check both
for damage.
Centrifuge door does not latch evenly
when closed
The basin position may need to be adjusted or the latch plate in the centrifuge
door may need to be aligned. See the Maintenance and Calibration chapter in
this manual for door alignment directions.
Centrifuge door does not stay
latched
Note:
This is a SAFETY ISSUE. Remove the device from service immediately.
1. Inspect the 4 door arms and the 8 shoulder screws used in mounting the
door arms. If one is unscrewed or broken, the door may not be able to
contain a major centrifuge failure.
2. Inspect the latch hooks on the frame and the latch holes in the latch plate for
damage or misalignment.
Door and Door Lock System Electrical Failure Symptoms
This section describes common electrical failure symptoms for the door and door lock systems.
Table 4-16: Possible door lock electrical failures
Symptom
Possible Causes
24 V over current STS value
Remove the door lock assembly and measure the solenoid coil resistance while
gently manipulating the assembly to try and find a short in the coil.
Door position Hall-effect sensor
does not report when door is closed
Inspect the south-pole magnet in the latch plate. It may be damaged or have
fallen out. A diagnostic tool can be made with the latch magnet spare.
Troubleshooting
4-20
Trima Accel
®
Automated Blood Collection System • Service Manual
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...