Plasma Only (Plasma Only Collect) Substate
This substate occurs after the Channel Setup substate in plasma/RBC procedures where plasma is
collected. The plasma valve opens and remains open until the plasma target volume is achieved. After
completion of this substate, the Trima Accel system advances to the RBC Collect substate if RBC is
chosen.
Table 3-34: Plasma Only Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
Qac
Qin
Qplas
Qplat
Qret
Return
Collect
Return
3000
Exit Condition
When the target (procedure-specific) volume is collected in the plasma bag.
PCA (Platelet Concentration Adjustment) Substate
This substate occurs after all plasma is collected during the Platelet Plasma substate. This substate only
occurs if the Trima Accel system calculates that the volume in the platelet bag will be lower than it should
be by a predetermined volume after the Mid Run substate. This usually occurs when the concurrent
plasma product collected is a small or medium product. During this substate the plasma valve moves to
the return position and adds plasma to the platelet bag to reach the needed volume. When the target
volume is added to the platelet bag, the Trima Accel system advances to the Mid Run substate.
Table 3-35: PCA Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
Qac
Qin
Qplas
Qplat
Qret
Collect
Return
Return
3000
Exit Condition
When the target (donor-specific) volume is collected in the platelet bag.
Software Description
3-28
Trima Accel
®
Automated Blood Collection System • Service Manual
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...