Blood Rinseback State
The Rinseback state returns the donor's blood. Variations include plasma or saline assisted rinseback.
The hematocrit in the RBC line is configured to maximize blood return. This system state ends when the
lower-level sensor detects air. Successful completion of the Blood Rinseback state results in the following
conditions and actions.
Exit Condition
Trima Accel System
Condition
Trima Accel System Action
All Blood Rinseback substates
are completed.
Under algorithm control.
The Trima Accel system advances to the Donor
Disconnect state.
Rinseback Substates
The Rinseback substates evacuate the channel and return remaining blood to the donor.
Rinseback Lower (Rinseback Lower Reservoir) Substate
All valves are commanded to the return position and the inlet, AC, plasma, and platelet pumps are
commanded to zero. The return pump runs until air is present at the lower-level sensor, or a
predetermined volume of return has been pumped. The Trima Accel system then advances to the
Rinseback Recirculate substate.
Table 3-185: Rinseback Lower Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
0
0
0
0
Qret
Return
Return
Return
Qcent
Exit Condition
When the lower-level sensor detects air or 10 mL processed by the return pump.
Rinseback Recirculate Substate
The inlet and return pumps are commanded to predetermined speeds to recirculate the blood in the
channel before returning it to the donor. After a predetermined inlet volume is processed, the Trima
Accel system advances to the Rinseback substate.
Table 3-186: Rinseback Recirculate Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
0
100
Qplas
Qplat
100
Return
Return
Return
Qcent
Exit Condition
When 50 mL is processed by the inlet pump.
Software Description
3-114
Trima Accel
®
Automated Blood Collection System • Service Manual
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...