Metered Storage Solution Disconnect State
The MSS Disconnect state checks that the operator has correctly sealed the product and additive
solution bags following the MSS state.
Upon completion of the MSS state, the Trima Accel system waits for the operator to confirm the
products and storage solution have been disconnected. After the operator confirms this information, the
system unloads the cassette and proceeds to the Post Run state.
Exit Condition
Trima Accel
System Condition
Trima Accel System Action
The operator pressed the
Continue
button.
Paused.
The Trima Accel system prompts the operator to remove the
disposable set. The Trima Accel system unloads the cassette
and proceeds to the Post Run state.
MSS Disconnect Substates
Upon completion of the MSS state, the Trima Accel system waits for the operator to confirm the
products and storage solution have been disconnected. After the operator confirms this information, the
system unloads the cassette and proceeds to the Post Run state.
Open Valves Substate
All valves are moved to their open positions. When the valve position sensors sense all three open
positions, the Trima Accel system advances to the next substate.
Table 3-216: Open Valves Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
0
0
0
0
0
Open
Open
Open
0
Exit Condition
When the valves are detected in the commanded position within 10 seconds.
Start Pumps Substate
All pumps are commanded to their unload speeds.
Table 3-217: Start Pumps Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
60
60
60
60
-110
Open
Open
Open
0
Exit Condition
When the pumps are detected at commanded speed within 10 seconds.
Software Description
3-126
Trima Accel
®
Automated Blood Collection System • Service Manual
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...