![Terumo BCT Trima Accel Скачать руководство пользователя страница 206](http://html1.mh-extra.com/html/terumo-bct/trima-accel/trima-accel_service-manual_1090436206.webp)
Stop Pumps Substate
The pumps are commanded to stop and the operator is prompted to press
Continue
. When the operator
presses
Continue
, the Trima Accel system advances to the Post Run state.
Table 3-109: Stop Pumps Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
0
0
0
0
0
Open
Open
Open
0
Metered Storage Solution State
The Metered Storage Solution (MSS) state administers the automatic addition of metered storage
solution.
This system state begins when the Trima Accel system determines if the run is a normal metered storage
run or a “salvage case.” This is determined by flags during the Rinseback protocols. A “salvage case” is
defined as a run where the machine does not return sufficient volume from the reservoir and channel
back to the donor. If the run is a “salvage case,” the machine will flag the final product for possible WBC
contamination. Successful completion of the Metered Storage Solution state results in the following
conditions and actions.
Exit Condition
Trima Accel System
Condition
Trima Accel System Action
All MSS substates are
completed.
Paused.
The Trima Accel system proceeds to the MSS
Disconnect state.
Metered Storage Solution Substates
Metered Storage Solution substates administer metered storage solution to the collected product.
MSS Setup (Metered Storage Solution Setup) Substate
This substate waits for the operator to confirm or cancel the automatic metered storage procedure. This
state transitions upon receiving message to continue from the GUI task.
Table 3-110: MSS Setup Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
0
0
0
0
0
Return
Return
Return
0
Exit Condition
When the operator touches the
Continue
button.
Version 6.0 Software Description
Trima Accel
®
Automated Blood Collection System • Service Manual
3-73
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...