Version 6.0.6 Dlog Information
This is the dlog header, periodic data, and episodic data that is relevant to the valve system.
The header is the function name and/or the unit of value for the periodic data.
The periodic data is device hardware values and device software information. This data is written into
the dlog every six seconds as long as the device is powered on. A row in the dlog with periodic data does
not have episodic data in it.
The episodic data is detailed and coded messages from the safety or control systems. These could be
values from hardware or software messages from the procedure. A row in the dlog with episodic data does
not have periodic data in it.
Header/Periodic Information
Table 4-6: Valve periodic information
Header Line 1
Header Line 2
Definition
CollectValveCmd
collect valve command
The commanded position of the collect valve.
Possible values are "O" (open); "C" (collecting); "R"
(returning); or "U" (unknown).
CollectValvePos
collect valve position
The measured position of the collect valve. Possible
values are "O" (open); "C" (collecting); "R" (returning);
or "U" (unknown).
PlasmaValveCmd
plasma valve command
The commanded position of the plasma valve.
Possible values are "O" (open); "C" (collecting); "R"
(returning); or "U" (unknown).
PlasmaValvePos
plasma valve position
The measured position of the plasma valve. Possible
values are "O" (open); "C" (collecting); "R" (returning);
or "U" (unknown).
RBCValveCmd
RBC valve command
The commanded position of the RBC valve. Possible
values are "O" (open); "C" (collecting); "R" (returning);
or "U" (unknown).
RBCValvePos
RBC valve position
The measured position of the RBC valve. Possible
values are "O" (open); "C" (collecting); "R" (returning);
or "U" (unknown).
RBC_1_AC
AC volume past the valve to RBC bag
1 (mL)
The current total amount of AC in the product inside
the RBC1 bag (mL).
RBC_1_Dose
RBC dose past the valve to RBC bag
1 (mL)
The RBC1 dose at 100% HCT. For example: If the HCT
of collected RBC product is configured at 80%, the
“RBC dose at 100% value” ÷ “RBC collect volume” =
80%.
Troubleshooting
4-10
Trima Accel
®
Automated Blood Collection System • Service Manual
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...