![Terumo BCT Trima Accel Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/terumo-bct/trima-accel/trima-accel_service-manual_1090436020.webp)
Boot Sequence
This section describes the boot sequence and when to interrupt that sequence for troubleshooting and
maintenance purposes.
When it is turned on, the Trima Accel device performs a low-level memory check, generating an alarm if
this check fails. If this check is successful, the device moves on to the boot sequence.
It is possible to interrupt the boot sequence for three different scenarios.
The first scenario is to enter Single-Step mode, which is used to connect the device to STS with an FTP
connection, usually after the safety computer fails to boot and dlogs need to be recovered. To enter
Single-Step mode, press and hold the pause button after the memory check until the screen shows that
the pause button is detected. Then press the pause button to advance through each step of the boot
sequence. Stop at the message line “press and release pause button to initialize serial port driver.”
In FTP mode, STS does not auto-discover the Trima Accel device. To connect, select device type
Trima
from the drop-down menu and enter the serial number, or enter the IP address 172.21.127.255 to
connect to a new hard drive. Accept any FTP warnings that appear.
The second scenario is to enable the software load installation script, which is used after new software has
been loaded and needs to be installed. To run the installation script, press and hold the pause and stop
buttons after the memory check, a few seconds after the power is on, and hold the buttons until the
message line “Installation script found—release buttons to execute.” Wait for another message line
reading “Installation complete—cycle power to restart.” Then boot to the two-button donor screen, and
check to make sure the changes took place.
Note:
The device allows only
one
chance to interrupt the normal boot sequence to install new files or software. All
calibration and software data is lost if the device boots to the two-button donor screen after uploading files to the
device. If this installation script is not activated on the first boot, then calibration and software transfer must be
performed again.
The third scenario is to enter Service mode, which is used for software installations, updates, device
maintenance, and device troubleshooting. To enter Service mode, press and hold the pause and stop
buttons after the screen shows the volume checks, and do not release the buttons until the screen
prompts you to do so.
After the boot sequence is finished, the device then checks the power supply voltages, performs valve
position tests, checks the leak detector voltage, and tests the door lock functionality. Without errors to
any of these items the software is then loaded into Procedure mode (two-button donor screen).
Boot Sequence
Trima Accel
®
Automated Blood Collection System • Service Manual
1-11
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...