![Terumo BCT Trima Accel Скачать руководство пользователя страница 201](http://html1.mh-extra.com/html/terumo-bct/trima-accel/trima-accel_service-manual_1090436201.webp)
Saline Bolus (Saline Bolus Recovery) Substate
This substate occurs whenever a saline bolus is administered. The platelet pump runs at a predetermined
speed until a predetermined volume of saline bolus is delivered. At the same time the return pump runs
at the same speed to keep the reservoir volume approximately even. After a predetermined volume of
saline has been pumped, the Trima Accel system resumes the draw or return cycle that it was previously
in.
Saline Prime (Replacement Fluid Priming Recovery) Substate
This substate occurs the first time replacement fluid is administered in an RBC/Plasma procedure. The
platelet pump runs at a predetermined speed until a predetermined volume of saline is pumped to prime
the replacement fluid line.
WBC Chamber Purge (WBC Chamber Purge Recovery) Substate
This substate occurs during any procedure where a spillover recovery occurs or a very large inlet volume
is processed. The plasma pump increases in speed and the platelet pump stops, purging volume from the
LRS chamber. This continues until a predetermined plasma volume is pumped through the LRS
chamber.
RBC Chamber Purge (RBC Chamber Purge Recovery) Substate
This substate occurs during any procedure where a platelet, plasma, RBC tubing set is loaded. This
recovery substate always occurs at the beginning of the Channel Setup substate. The plasma pump runs
at a predetermined speed and the platelet pump stops, purging volume from the LRS chamber. This
continues until a predetermined plasma volume is pumped.
Plasma Valve Motion (Plasma Valve Move Recovery) Substate
This substate occurs whenever the plasma valve changes position during a platelet collection procedure.
This recovery stops the pumps for one second while the valve moves to prevent fluid from entering the
plasma bag while the valve moves through the open position.
Settle Channel Substate
This substate occurs only after the WBC Chamber Purge substate. This substate settles the cells in the
channel prior to resuming collection. During this recovery the inlet pump flow rate decreases to less than
a predetermined amount and pumps a predetermined inlet volume.
Clear Line Substate
This substate occurs after WBC Purge recoveries. It functions to clear RBC/WBC out of the platelet
line. The inlet, platelet, and plasma pumps are limited to predetermined speeds. This substate ends when
a predetermined platelet volume has been pumped.
Software Description
3-68
Trima Accel
®
Automated Blood Collection System • Service Manual
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...