![Terumo BCT Trima Accel Скачать руководство пользователя страница 179](http://html1.mh-extra.com/html/terumo-bct/trima-accel/trima-accel_service-manual_1090436179.webp)
Successful completion of the Disposable Tests state results in the following conditions and actions.
Exit Condition
Trima Accel System
Condition
Trima Accel System Action
All disposable tests have
passed.
Pause.
The Trima Accel system prompts the operator to
connect the AC line.
Trima Accel version 6.0 introduces new disposables that require a different set of disposable tests than
previous software versions. During the Pressurize Inlet Line substate, the system determines which type
of disposable is loaded on the system based on the pressure measured at the APS. If the APS reads less
than 300 mmHg the set is recognized as a “RAS” set and the new disposable test tree is used. If the
pressure is greater than 500 mmHg, the set is recognized as a “non-RAS” set and the original disposable
tree is used.
Software Description
3-46
Trima Accel
®
Automated Blood Collection System • Service Manual
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...