MSSPFR (Metered Storage Solution Power Fail Recovery) Substate
This substate recovers from a power failure. It will branch to the last substate being executed at the time
of the failure. This substate generates no alarms or alerts.
Table 3-195: MSSPFR Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
0
0
0
0
0
Return
Return
Return
0
Verify Closure (Verify Channel Clamps are Closed) Substate
This substate verifies that the channel clamps are closed. It runs the pumps to see if fluid can be pulled
from the channel. If fluid is pulled from the channel into the reservoir, a channel clamp alert is generated
and the product is flagged for WBC contamination. This alert is generated a maximum of two times in
this substate. The operator is then allowed to continue after the second alert. This substate is not
executed if it is a salvage case MSS run.
Table 3-196: Verify Closure Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
10, 0, 30
0, 20
0
0
0
Return
Return
Return
0
Exit Condition
When fluid is not detected by the AC detector and APS is greater than 250 mmHg.
Reservoir Setup Substate
This substate empties the reservoir. The return pump runs until the lower-level sensor detects air. This
state is not executed if this is determined to be a salvage case MSS run. (Reservoir Setup_2 is identical to
Reservoir Setup except for the name).
Table 3-197: Reservoir Setup Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
0
27, 0
0
0
27, 0
Return
Return
Return
0
Exit Condition
When the lower-level sensor detects air.
Version 6.1 Software Description
Trima Accel
®
Automated Blood Collection System • Service Manual
3-119
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...