plasma product collected is a small or medium product. During this substate the plasma valve moves to
the return position and adds plasma to the platelet bag to reach the needed volume. When the target
volume is added to the platelet bag, the Trima Accel system advances to the Mid Run substate.
Table 3-92: PCA Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
Qac
Qin
Qplas
Qplat
Qret
Collect
Return
Return
3000
Exit Condition
When the target (donor-specific) volume is collected in the platelet bag.
Mid Run (Platelet Only Collection) Substate
This substate occurs after the Platelet Plasma substate (and the Platelet Concentration Adjustment
substate, if it occurs) to collect the remainder of the platelet product needed to reach the target yield.
This substate only occurs if more collected platelets are needed after the Concurrent Platelet Plasma
substate. During this substate the plasma pump runs slowly.
Table 3-93: Mid Run Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
Qac
Qin
Qplas
Qplat
Qret
Collect
Return
Return
3000
Exit Condition
When the target (donor-specific) volume is collected in the platelet bag.
PIR (Platelet Inventory Recovery) Substate
This substate occurs after the Mid Run substate if the collect flow is less than a predetermined rate during
Platelet Only Collection. Otherwise, the Trima Accel system skips this substate and advances to the
Extended PCA substate. During Platelet Inventory Recovery the platelet pump ramps up in speed to its
maximum value and continues running until predetermined platelet volume has been pumped. The
Trima Accel system advances to the Extended PCA substate.
Table 3-94: PIR Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
Qac
Qin
Qplas
Qplat
Qret
Collect
Return
Return
3000
Exit Condition
When the target (donor-specific) volume is collected in the platelet bag.
Software Description
3-64
Trima Accel
®
Automated Blood Collection System • Service Manual
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...