Table 38:
Percent-Good Data Format
Hex Digit
Binary
Byte
Field
Description
1-4
1,2
ID Code
Stores the percent-good data identification word (MSB=04h
LSB=00h).
5,6
3
Depth cell 1, Field
1
Stores percent-good data for depth cell #1, field 1. See above.
7,8
4
Depth cell 1, Field
2
Stores percent-good data for depth cell #1, field 2. See above.
9,10
5
Depth cell 1, Field
3
Stores percent-good data for depth cell #1, field 3. See above.
11,12
6
Depth cell 1, Field
4
Stores percent-good data for depth cell #1, field 4. See above.
13-1028
7-514
Depth cell 2 – 128
(if used)
These fields store percent-good data for depth cells 2 through 128
(depending on
), following the same
format as listed above for depth cell 1.
These fields contain information about the status and quality of DVL data. A value of 0 means the meas-
urement was good. A value of 1 means the measurement was bad.
Table 39:
Status Data Format
Hex Digit
Binary
Byte
Field
Description
1-4
1,2
ID Code
Stores the status data identification word (00 05h).
5,6
3
Depth cell 1, Field
1
Stores status data for depth cell #1, beam #1. See above.
7,8
4
Depth cell 1, Field
2
Stores status data for depth cell #1, beam #2. See above.
9,10
5
Depth cell 1, Field
3
Stores status data for depth cell #1, beam #3. See above.
11,12
6
Depth cell 1, Field
4
Stores status data for depth cell #1, beam #4. See above.
13-1028
7-514
Depth cell 2 – 128
(if used)
These fields store status data for depth cells 2 through 128 (de-
pending on the
) for all four beams.
These fields follow the same format as listed above for depth cell 1.