Math and Measure
To remove a connection between process blocks, touch the line. A scissor icon appears above it.
Confirm whether or not to "cut" the connection.
Add Other Terminals to Processing Web
Other measurements or math functions can share the same processing web. For example, you may
wish to create a math function of the waveform that would result from a sub-processing node, while
the final output of the full process is a measurement parameter.
To add a terminal:
1. Open the Math or Measure dialog behind the Web Editor dialog.
2. Touch the Web Edit icon following the location (P
x
or F
x
). The summary changes to Web Edit,
and a new terminal block appears on the Web Editor peg board.
3. Connect the output pin of the desired process to the input pin of the terminal. You can select
outputs that are already pinned to other blocks.