Model 200E Instruction Manual
Troubleshooting & Repair
path and vacuum, most likely through the Perma Pure dryer flow paths. See
troubleshooting the Perma Pure dryer later in this chapter.
•
If the sample and vacuum pressures are around their nominal values (28 and
<10 in-Hg-A, respectively) and the flow still displays
XXXX
, carry out a leak
check as described in Section 11.5.2
•
If gas flows through the instrument during the above tests but goes to zero or is low
when it is connected to zero air or span gas, the flow problem is not internal to the
analyzer but likely caused by the gas source such as calibrators/generators, empty
gas tanks, clogged valves, regulators and gas lines.
•
If an IZS or Zero/Span valve option is installed in the instrument, press
CALZ
and
CALS
. If the sample flow increases, suspect a bad Sample/Cal valve.
•
If none of these suggestions help, carry out a detailed leak check of the analyzer as
11.2.1.2. Ozone Flow is Zero or Low
If there is zero or a low (<50 cm³/min) ozone flow, the unit displays an
OZONE FLOW
WARNING
message on the front panel and a value between 0.0 and 50 cm³/min for the
actual ozone flow as measured by the internal mass flow meter. In this case, carry out the
following steps:
•
Check the actual flow rate through the ozone dryer by attaching the outlet of an
external flow meter to the inlet port of the dryer. This inlet port is inside the analyzer
at the end of the plastic particle filter (Section 9.3.2 for illustration). If there is
nominal flow (about 140 cm³/min from 80 cm³/min O
3
flow and 60 cm³/min purge
flow), consult customer service as there is a problem with the firmware or
electronics.
•
If the actual flow is low or zero, check if the pump operates properly. The
RCEL
pressure should be below 10 in-Hg-A at sea level. If it is above 10”, rebuild the
pump (Section 9.3.4). Check the spare parts list in Appendix B on how to order
pump rebuild kits.
•
Check if the particle filter is clogged.
Briefly
remove the particle filter to see if this
improves the flow. Be very cautious about handling the Perma Pure dryer fittings -
refer to Section 9.3.2 on proper handling instructions. If the filter is clogged, replace
it with a new unit. If taking off this filter does not solve the problem, continue to the
next step. Do not leave the Perma Pure dryer without filter for more than a few
seconds, as you may draw in dust, which will reduce the performance of the dryer.
•
A leak between the flow meter and the reaction cell (where the flow-determining
critical orifice is located) may cause a low flow (the system draws in ambient air
through a leak after the flow meter). Check for leaks as described in Section 11.5.1.
Repair the leaking fitting, line or valve and re-check.
•
The most likely cause for zero or low ozone flow is a clogged critical flow orifice or
sintered filter within the orifice assembly. The orifice that sets the ozone flow is
located on the reaction cell. Check the actual ozone flow by disconnecting the tube
from the reaction cell and measuring the flow going into the cell. If this flow is
correct (~80 cm³/min), the orifice works properly. If this flow is low, replace or
clean the orifice. The orifice holder assembly allows a quick and easy replacement of
the orifice, refer to Section 9.3.10 on how to do this. Appendix B lists a spare part kit
with a complete orifice assembly that allows a quick replacement with minimum
044100102 Rev A
197
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...