Model 200E Instruction Manual
Theory of Operation
The M200E utilizes a dual-dielectric design. This method utilizes a glass tube with hollow
walls. The outermost and innermost surfaces are coated with electrically conductive
material. The air flows through the glass tube, between the two conductive coatings, in
effect creating a capacitor with the air and glass acting as the dielectric. The layers of glass
also separate the conductive surfaces from the air stream to prevent reaction with the O
3
.
As the capacitor charges and discharges, electrons are created and accelerated across the
air gap and collide with the O
2
molecules in the air stream splitting them into elemental
oxygen. Some of these oxygen atoms recombine with O
2
to O
3
.
The quantity of ozone produced is dependent on factors such as the voltage and frequency
of the alternating current applied to the CD cells. When enough high-energy electrons are
produced to ionize the O
2
molecules, a light emitting, gaseous plasma is formed, which is
commonly referred to as a corona, hence the name corona discharge generator.
10.2.6. Perma Pure
®
Dryer
The air supplied to the O
3
generation system needs to be as dry as possible. Normal room
air contains a certain amount of water vapor, which greatly diminishes the yield of ozone
produced by the ozone generator. Also, water can react with other chemicals inside the O
3
Generator to produce chemicals that damage the optical filter located in the reaction cell
(Table 10-4) such as ammonium sulfate or highly corrosive nitric acid.
To accomplish this task the M200E uses a Perma Pure
®
single tube permeation dryer. The
dryer consists of a single tube of Nafion
®
, a co-polymer similar to Teflon
®
that absorbs
water very well but not other chemicals. The Nafion
®
tube is mounted within an outer,
flexible plastic tube. As gas flows through the inner Nafion
®
tube, water vapor is absorbed
into the membrane walls. The absorbed water is transported through the membrane wall
and evaporates into the dry, purge gas flowing through the outer tube, countercurrent to
the gas in the inner tube (Figure 10-7).
Figure 10-7: Semi-Permeable Membrane Drying Process
This process is called per-evaporation
and
is driven by the humidity gradient between the
inner and outer tubes as well as the flow rates and pressure difference between inner and
outer tubing. Unlike micro-porous membrane permeation, which transfers water through a
relatively slow diffusion process, per-evaporation is a simple kinetic reaction. Therefore, the
drying process occurs quickly, typically within milliseconds. The first step in this process is a
chemical reaction between the molecules of the Nafion
®
material and water, other chemical
044100102 Rev A
165
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...