Model 200E Instruction Manual Instrument
Maintenance
•
To fulfill CE Mark safety requirements, the front panel locking screw must be
installed at all times during operation of the analyzer.
•
Re-start the analyzer.
9.3.2. Changing the O
3
Dryer Particulate Filter
The air for the O
3
generator passes through a Perma Pure
©
dryer, which is equipped with a
small particulate filter at its inlet. This filter prevents dust from entering the Perma Pure
©
dryer and degrading the dryer’s performance over time. To change the filter according to
the service interval in Table 9-1:
•
Before starting the procedure, check and write down the average RCEL pressure and
the OZONE flow values.
•
Turn off the analyzer, unplug the power cord and remove the cover.
•
Unscrew the nut around the port of the filter using 5/8” and 9/16” wrenches and by
holding the actual fitting body steady with a 7/16” wrench.
CAUTION
Risk of significant leak. Make sure to use proper wrenches and to not turn the
fitting against the Perma Pure
©
dryer. This may loosen the inner tubing and cause
large leaks.
•
Take off the old filter element and replace it with a suitable equivalent
(TAPI part# FL-3).
Figure 9-2: Particle Filter on O
3
Supply Air Dryer
•
Holding the fitting steady with a 5/8” wrench, tighten the nut with your hands. If
necessary use a second wrench but do not over-tighten the nut.
•
Replace the cover, plug in the power cord and restart the analyzer.
•
Check the O
3
flow rate, it should be around 80 cm³/min ± 15. Check the RCEL
pressure, it should be the same value as before.
044100102 Rev A
145
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...