Model 200E Instruction Manual
APPENDIX A
Setup Variable
Numeric
Units
Default
Value
Value
Range
Description
NO_SPAN2
Conc.
400
4-20000
Target NO concentration during
span calibration of range 2.
NO2_SPAN2 Conc.
400 4-20000
Target
NO
2
concentration during
converter efficiency calibration of
range 2.
NOX_SLOPE2 PPB/mV
3,5
PPM/mV
4
1 0.25-4
NO
X
slope for range 2.
NOX_OFFSET2 mV
0
-10000-
10000
NO
X
offset for range 2.
NO_SLOPE2 PPB/mV
3,5
PPM/mV
4
1
0.25-4
NO slope for range 2.
NO_OFFSET2 mV 0
-10000-
10000
NO offset for range 2.
RANGE_MODE —
SNGL SNGL,
IND,
AUTO
Range control mode. Enclose value
in double quotes (") when setting
from the RS-232 interface.
PHYS_RANGE1
PPM
2
0.1-2500 Low pre-amp range.
PHYS_RANGE2
PPM
20
0.1-2500 High pre-amp range.
CONC_RANGE1 Conc. 500
10-
20000
D/A concentration range 1 or
range for NO
X
.
CONC_RANGE2
1
Conc.
500
10-
20000
D/A concentration range 2 or
range for NO.
CONC_RANGE3
1
Conc.
500
10-
20000
D/A concentration range 3 or
range for NO
2
.
RCELL_SET ºC 50
Warnings:
45-55
30-70
Reaction cell temperature set point
and warning limits.
MANIFOLD_SET
1
ºC
50
Warnings:
45-55
30-70
Manifold temperature set point and
warning limits (if installed).
CONV_TYPE — MOLY
NONE,
MOLY,
CONV,
O3KL
Converter type. “CONV” is mini-
hicon, O3KL is O
3
killer only (no
converter installed). Enclose value
in double quotes (") when setting
from the RS-232 interface.
CONV_SET ºC 315
Warnings:
305-325
0-800
Converter temperature set point
and warning limits for moly
converter.
CONV_TEMP_TRIG
Cycles
10
0-100
Number of converter temperature
errors required to trigger warning.
BOX_SET ºC
30
Warnings:
7-48
0-70
Nominal box temperature set point
and warning limits.
PMT_SET
ºC
7 Warnings:
5-12
0-40
PMT temperature warning limits.
Set point is not used.
SFLOW_SET cc/m
500
Warnings:
350–600
0–1000
Sample flow warning limits. Set
point is not used.
SAMP_FLOW_SLOPE —
1
0.001-
100
Slope term to correct sample flow
rate.
044100110 Rev A
235
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...