Operating Instructions
Model 200E Instruction Manual
communicate with your local area network. To perform this configuration, you will need to
get the following information from your network administrator:
•
Gateway IP address, typically a string of numbers with four packets of 1 to 3
numbers each:
nnn.nnn.nnn.nnn
, for example:
192.168.76.1
.
•
Instrument IP address, typically a string of numbers very similar to the Gateway IP
address, for example:
192.168.76.55
. Note that this instrument IP address must
be a static address and cannot be a dynamic IP address. DHCP and analyzer host
names are not supported at this time.
•
Subnet Mask, typically a string of numbers such as
255.255.252.0
.
•
The communications port number is set to 3000 by default and should not be
changed.
The following procedures assume that the Ethernet card is already installed in your
analyzer. If you are retro-fitting the analyzer with an Ethernet Option 63, please make sure
to enable the internet option as described in Section 6.9.6 and set the baud rate to 115000
as described in Section 6.9.7. Refer to the installation instructions of that option for more
information.
Once Ethernet is enabled for the COM2 port, the first step in configuring the Ethernet option
is to set the card from pass-through mode to configuration mode, so that TCP/IP parame-
ters can be edited.
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS DIAG
EXIT
SAMPLE RANGE = 500.0 PPB O3 =XXX.X
< TST TST > CAL
SETUP
ETHERNET STOPPED
SETUP X.X
COMMUNICATIONS MENU
ID
INET
COM1
COM2
EXIT
INTERNET CONFIGURATION
GTWY IP SNET PORT STRT
STOP
EXIT
STOPPING ETHERNET
If the message
ETHERNET NOT
STOPPED
appears here, the
Ethernet card either was already
in configuration mode or might
not be responding.
Step two in the configuration process is to edit the TCP/IP parameters.
82
044100102 Rev A
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...