Model 200E Instruction Manual
Operating Instructions
Table 6-2: Test Functions Defined
Display Parameter
Units
Description
RANGE RANGE
PPB, PPM,
UGM, MGM
The full scale limit to which the analyzer’s analog outputs are
currently set. This is not the physical range of the instrument. If
the
Auto
range mode has been selected, two
RANGE
functions
will appear. If
the
IND
(independent) range mode has been
selected, three
RANGE
functions will appear:
NOX STB
STABILITY
PPB, PPM,
UGM, MGM
The stability is a standard deviation of the NO
X
concentration
over 25 samples, each recorded every 10 seconds. A low
nox
stb
value indicates low variability in NO
X
.
SAMP FLW SAMPLE FLOW
cm³/min
(cc/m)
The flow rate of the sample gas through the reaction cell. This
value is not measured but calculated from the sample pressure.
OZONE FL
OZONE FLOW
cm³/min
(cc/m)
Flow rate of the O
3
gas stream as measured with a flow meter
PMT
PMT Signal
MV
The raw output voltage of the PMT.
NORM
PMT
NORMALIZED
PMT Signal
MV
The output voltage of the PMT after normalization for auto-zero
offset and temperature/pressure compensation (if activated).
AZERO AUTO-ZERO MV
The PMT signal with zero NO
X
,
which is usually slightly different
from 0 V. This offset is subtracted from the PMT signal and
adjusts for variations in the zero signal.
HVPS
HVPS
V
The PMT high voltage power supply.
RCELL
TEMP
REACTION CELL
TEMP
°
C
The current temperature of the reaction cell.
BOX TEMP
BOX
TEMPERATURE
°
C
The ambient temperature of the inside of the analyzer case.
PMT TEMP
PMT
TEMPERATURE
°
C
The current temperature of the PMT.
IZS TEMP
IZS
TEMPERATURE
1
°
C
The current temperature of the internal zero/span option. Only
appears when IZS option is enabled.
MOLY
TEMP
CONV
TEMPERATURE
°
C
The current temperature of the NO
2
converter.
RCEL
REACTION CELL
PRESSURE
in-Hg-A
The current gas pressure of the reaction cell as measured at the
vacuum manifold. This is the vacuum pressure created by the
external pump.
SAMP
SAMPLE
PRESSURE
in-Hg-A
The current pressure of the sample gas as it enters the reaction
cell, measured between the NO/NO
x
and Auto-Zero valves.
NOX
SLOPE
NO
x
SLOPE
- -
The slope of the current NO
x
calibration as calculated from a
linear fit during the analyzer’s last zero/span calibration.
NOX OFFS NO
x
OFFSET
MV
The offset of the current NO
x
calibration as calculated from a
linear fit during the analyzer’s last zero/span calibration.
NO SLOPE NO SLOPE
- -
The slope of the current NO calibration as calculated from a
linear fit during the analyzer’s last zero/span calibration.
NO OFFS
NO OFFSET
MV
The offset of the current NO calibration as calculated from a
linear fit during the analyzer’s last zero/span calibration.
NO2
NO
2
concentration
PPB, PPM,
UGM, MGM
The current NO
2
concentration in the chosen unit.
NOX
NO
x
concentration
PPB, PPM,
UGM, MGM
The current NO
x
concentration in the chosen unit.
NO NO
concentration
PPB, PPM,
UGM, MGM
The current NO concentration in the chosen unit.
TEST TEST
SIGNAL
2
MV
Signal of a user-defined test function on output channel A4.
TIME
CLOCK TIME
hh:mm:ss
The current day time for iDAS records and calibration events.
044100102 Rev A
43
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...