Table 6-20: iDAS Data Channel Properties
Property Description
Default
Setting
Range
NAME
The name of the data channel.
“NONE”
Up to 6 letters and
digits (more with
APICOM, but only
the first six are
displayed on the
front panel).
TRIGGERING
EVENT
The event that triggers the data channel to
measure and store its data parameters. See
APPENDIX A-5 for a list of available
triggering events.
ATIMER
Any allowed event.
NUMBER AND
LIST OF
PARAMETERS
A User-configurable list of data types to be
recorded in any given channel. See
APPENDIX A-5 for a list of available
parameters.
PMTDET Any
available
concentration,
temperature,
pneumatic or
diagnostic
parameter.
STARTING DATE
The starting date when a channel starts
collecting fata
01-JAN-03 Any actual date in
the past or future.
SAMPLE PERIOD The amount of time between each data point
that is averaged into one mean reported
every REPORT PERIOD.
000:01:00 000:00:01
to
366:23:59
(Days:Hours:Minut
es)
REPORT PERIOD
The amount of time between each channel
data point.
000:01:00 000:00:01
to
366:23:59
(Days:Hours:Minut
es)
NUMBER OF
RECORDS
The number of reports that will be stored in
the data file. Once the specified limit has
been exceeded, the oldest data are over-
written to make space for new data.
100
1 to 1 million,
limited by available
storage space.
RS-232 REPORT
Enables the analyzer to automatically report
channel values to the RS-232 ports.
OFF
OFF or ON
COMPACT
REPORT
Enables a compact version of the RS-232
reporting with up to five parameters sharing
the same output line.
OFF
OFF or ON
CHANNEL
ENABLED
Enables or disables the channel. Provides a
convenient means to temporarily disable a
data channel.
ON
OFF or ON
CAL HOLD OFF
Disables sampling of data parameters while
instrument is in calibration mode.
Note that - when enabled here - there is
also a length of the DAS HOLD OFF after
calibration mode, which is set in the VARS
menu. (Section 6.5.)
OFF
OFF or ON
6.10.1.2. iDAS Parameters
Data parameters are types of data that may be measured and stored by the iDAS. For each
T-API analyzer model, the list of available data parameters is different, fully defined and not
customizable. Appendix A-5 lists firmware specific data parameters for the M200E. The
044100102 Rev A
87
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...