Troubleshooting & Repair
Model 200E Instruction Manual
and sub-assemblies in the analyzer. See the wiring interconnect diagram (document
04504) and interconnect list (document 04496) in Appendix D.
11.1.1. Warning Messages
The most common and/or serious instrument failures will result in a warning message
displayed on the front panel. Table A-2 in Appendix A-3 contains a list of warning
messages, along with their meaning and recommended corrective action.
The analyzer will alert the user that a warning is active by displaying the keypad labels
MSG
and
CLR
on the front panel and a text message in the top center line of the display as
shown in this example:
SAMPLE AZERO WARNING NOX =123.4
< TST TST > CAL
MSG CLR SETUP
The analyzer will also issue a message to the serial port and cause the red FAULT LED on
the front panel to blink.
To view or clear a warning messages press:
SAMPLE
SYSTEM RESET
NOX = XXX.X
TEST
CAL
MSG
CLR
SETUP
If warning messages re-appear,
the cause needs to be found. Do
not repeatedly clear warnings
without corrective action.
Press
CLR
to clear the current
warning message.
If more than one warning is
active, the next message will
take its place.
Once the last warning has been
cleared, the analyzer returns to
SAMPLE
Mode.
SAMPLE
RANGE=500 PPB
NOX = XXX.X
< TST TST > CAL
MSG
CLR
SETUP
SAMPLE
SYSTEM RESET
NOX = XXX.X
< TST TST > CAL
MSG
CLR
SETUP
<TST TST
> keys replaced with
TEST
key. Pressing
TEST
deactivates warning messages
until new warning(s) are activated.
MSG
indicates that warning
messages are active.
All Warning messages are hidden,
but
MSG
button appears
Figure 11-1: Viewing and Clearing Warning Messages
11.1.2. Fault Diagnosis with Test Functions
Besides being useful as predictive diagnostic tools, the TEST functions, viewable from the
front panel, can be used to isolate and identify many operational problems when combined
with a thorough understanding of the analyzer’s theory of operation (Chapter 10). We
recommend to use the APICOM remote control program to download, graph and archive
TEST data for analysis and long-term monitoring of diagnostic data (
).
Section 6.11.5
190
044100102 Rev A
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...