Calibration Procedures
Model 200E Instruction Manual
gas inlets rather than through the sample inlet and that the zero and span operations are
initiated directly and independently with dedicated keys
CALZ and CALS
.
Connect the sources of zero air and span gas to the respective ports on the rear panel
(Figure 3-2) to follow the setup in Figure 7-4. Set the expected NO and NO
X
values
according to the procedure described in Section 7.2.
Perform the calibration or calibration check according to the following flow chart:
Wait until
NOX
STB
falls below
0.05 ppb. This
may take several
minutes.
ZERO CAL M
NOX STB=XXX.X PPB
NOX=X.XX
< TST TST >
ZERO
CONC
EXIT
returns to the
main
SAMPLE
display
ZERO CAL M NOX STB=XXX.X PPB NOX=X.XX
< TST TST > ENTR CONC
EXIT
Pressing
ENTR
changes the calibration of the instrument.
If either the
ZERO
or
SPAN
button fails to
appear, see Chapter 11
for troubleshooting tips.
ANALYZER ENTERS
ZERO CAL MODE.
EXIT
returns the unit to
SAMPLE
mode without
changing the calibration
values.
ZERO CAL M NOX STB=XXX.X PPB NOX=X.X
< TST TST > ZERO CONC
EXIT
ZERO CAL M NOX STB=XXX.X PPB NOX=X.X
X
< TST TST > CAL
CALZ
CALS
SETUP
SPAN CAL M
NOX STB=XXX.X PPB NOX=X.XX
< TST TST >
SPAN
CONC
EXIT
SPAN CAL M NOX STB=XXX.X PPB NOX=X.XX
< TST TST >
ENTR
CONC
EXIT
EXIT
returns to the
SAMPLE
mode without
changing the calibration
values.
SPAN CAL M NOX STB=XXX.X PPB NOX=X.XX
< TST TST >
SPAN
CONC
EXIT
ACTION:
Allow zero gas to enter the sample port at the
rear of the instrument.
SAMPLE RANGE = 500.0 PPB NOX= X.XX
< TST
TST >
CAL
CALZ
CALS
SETUP
SAMPLE
NOX STB=XXX.X PPB
NO = X.XX
< TST TST > CAL
CALZ
CALS
SETUP
Scroll the display to the
NOX
STB
test function. This function
calculates the stability of the
NO/NO
X
measurements.
The value of
NOX
STB
may jump
significantly. Wait
until it falls below
0.5
ppb. This may take
several minutes.
Pressing
ENTR
changes the calibration of the instrument.
Analyzer enters
SPAN
CAL
Mode.
116
044100102 Rev A
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...