Theory of Operation
Model 200E Instruction Manual
See Section 3.1 for information on connecting and operating the external pump.
Figure 10-5: External Pump Pack
Finally, the M200E requires a steady, high underpressure, which cannot be achieved reliably
over extended periods of time with small vacuum pumps. The external pump used for the
M200E has a very long lifetime and duty cycle and provides a very good vacuum for its
entire lifetime. However, the pump is too large to fit into the chassis of the analyzer.
10.2.2. Sample Gas Flow
The sample gas is the most critical flow path in the analyzer, as the medium has to be
routed through a variety of valves and tubes for the measurement of zero offset and
concentrations of both NO and NO
X
(and possibly the drying of the gas if the optional
sample dryer is installed). At any point before and in the reaction cell, the integrity of the
sample gas cannot be compromised.
Sample gas flow in the M200E analyzer is not a directly measured value, but is rather
calculated from the sample pressure using the flow principle across a critical orifice. In
general, the differential pressure ratio between sample pressure and reaction cell pressure
needs to exceed 2:1 to allow critical flow. The actual flow rate is then only dependent on
the size of the orifice and the upstream pressure. Refer to Section 10.2.9 for a detailed
description of critical flow.
The M200E uses a 0.010” (10 mil) diameter orifice to achieve a flow rate of about
450-500 cm³/min. With nominal pressures of 28 and 4 in-Hg-A for the sample and reaction
cell pressures, respectively. The necessary ratio of sample to reaction cell pressure of 2:1 is
largely exceeded and accommodates a wide range of possible variability in atmospheric
pressure and pump degradation. Once the pump degrades and vacuum pressure exceeds
14 in-Hg-A, the ratio between sample and vacuum pressures is less than 2:1 and, hence, a
critical flow rate can no longer be calculated. At this point, the instrument will display an
invalid sample flow rate measurement (
XXXX
).
162
044100102 Rev A
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...