CAUTION
The thermo-electric cooler needs to be mounted flat to the heat sink. If there is
any significant gap, the TEC might burn out. Make sure to apply heat sink paste
before mounting it and tighten the screws evenly and cross-wise.
•
Re-insert the TEC subassembly in reverse order. Make sure that the O-ring is placed
properly and the assembly is tightened evenly.
•
Re-insert the PMT/HVPS subassembly in reverse order and don’t forget the gasket
between HVPS and PMT. Use new plastic screws to mount the PMT assembly on the
PMT cold block.
•
Insert the LED and thermistor into the cold block, insert new drying packages and
carefully replace the end plate by making sure that the O-ring is properly in place.
Improperly placed O-rings will cause leaks, which – in turn – cause moisture to
condense on the inside of the cooler and likely cause a short in the HVPS.
•
Reconnect the cables and the reaction cell (evenly tighten these screws), replace the
sensor assembly into the chassis and fasten with four screws and washers.
•
Reconnect all electrical and pneumatic connections, leak check the system and
power up the analyzer.
•
Verify the basic operation of the analyzer using the ETEST and OTEST features or
zero and span gases, then carry out a hardware calibration of the analyzer
(Section 11.6.5) followed by a software calibration.
11.7. Technical Assistance
If this manual and its trouble-shooting / repair sections do not solve your problems,
technical assistance may be obtained from Teledyne-API, Customer Service, 6565 Nancy
Ridge Drive, San Diego, CA 92121. Phone: +1 858 657 9800 or 1-800 324 5190. Fax: +1
858 657 9816. Email: [email protected]. Before you contact customer
service, fill out the problem report form in Appendix C, which is also available online for
electronic submission at http://www.teledyne-api.com/forms/.
044100102 Rev A
223
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...