Table 11-2: Relay Board Status LEDs
LED
Color
Function
Fault Status
Indicated Failure(s)
LED Row 1 (center of board)
D1
red
Watchdog Circuit; I
2
C bus
operation.
Continuously
ON or OFF
Failed or halted CPU; faulty
motherboard, keyboard, relay
board; wiring between mother-
board, keyboard or relay board; +5
V power supply
D2
yellow Relay 0 - reaction cell
heater
Continuously
ON or OFF
Heater broken, thermistor broken
D3
yellow Relay 1 - NO
2
converter
heater
Continuously
ON or OFF
Heater broken, thermocouple
broken
D4
1
yellow Relay 2 - manifold heater
Continuously
ON or OFF
Heater broken, thermistor broken
D7
2
green
Valve 0 - zero/span valve
status
Continuously
ON or OFF
Valve broken or stuck, valve driver
chip broken
D8
2
green
Valve 1 - sample/cal valve
status
Continuously
ON or OFF
Valve broken or stuck, valve driver
chip broken
D9
green
Valve 2 - auto-zero valve
status
Continuously
ON or OFF
Valve broken or stuck, valve driver
chip broken
D10
green
Valve 3 - NO/NO
x
valve
status
Continuously
ON or OFF
Valve broken or stuck, valve driver
chip broken
LED Row 2 (top of board)
D5
yellow Relay 3 - IZS heater
Continuously
ON or OFF
Heater broken, thermistor broken
D6
yellow Relay 4 – (O
2
sensor heater
200EH/EM)
N/A N/A
D11
green
Valve 4 – Spare
N/A
N/A
D12
green
Valve 5 – Spare
N/A
N/A
D13
green
Valve 6 – Spare
N/A
N/A
D14
green
Valve 7 – Spare
N/A
N/A
D15 green
Mosfet1/DC
driver-Unused
N/A
N/A
D16
green
Mosfet2/DC driver -Unused N/A
N/A
1
Special configurations only
2
Only active for instruments with Z/S valve or IZS options installed
11.2. Gas Flow Problems
The M200E has two main flow paths, the sample flow and the flow of the ozone supply air.
With IZS or zero/span valve option installed, there is a third (zero air) and a fourth (span
gas) flow path, but either one of those is only controlled by critical flow orifices and not
displayed on the front panel or stored to the iDAS. The full flow diagrams of the standard
configuration and with options installed (Appendix D, document 04574) help in trouble-
shooting flow problems. In general, flow problems can be divided into three categories:
•
Flow is too high
•
Flow is greater than zero, but is too low, and/or unstable
044100102 Rev A
195
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...