Model 200E Instruction Manual
Operating Instructions
Press
SET>
to select the analog output channel to be configured:
DISPLAYED
AS
=
CHANNEL
CONC_OUT_1
=
A1
CONC_OUT_2
=
A2
CONC_OUT_3
=
A3
TEST
OUTPUT
=
A4
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AOUTS CALIBRATED:
NO
< SET
SET>
CAL
EXIT
FROM ANALOG I/O CONFIGURATION MENU
DIAG AIO
CONC_OUT_1 :5V, NO CAL
< SET SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_1 CALIBRATED: YES
< SET
CAL
EXIT
DIAG AIO
CONC_OUT_1 RANGE: 5V
SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2 CALIBRATED: NO
< SET
CAL
EXIT
DIAG AIO
CONC_OUT_1 VOLT–Z : 0 mV
U100 UP10 UP DOWN DN10 D100
ENTR EXIT
These keys increase / decrease the analog
output by 100, 10 or 1 counts. Continue
adjustments until the voltage measured at the
output of the analyzer and/or the input of the
recording device matches the value in the upper
right hand corner of the display to the tolerance
listed in Table 6-10.
The concentration display
will not change. Only
the voltage reading of your voltmeter will change.
DIAG AIO
CONC_OUT_1 VOLT–S : 4500 mV
U100 UP10 UP DOWN DN10 D100
ENTR EXIT
DIAG AIO
CONC_OUT_1 REC OFS: 0 mV
< SET
SET>
EDIT
EXIT
If AutoCal is ON, go to
Section 6.7.3
EXIT
ignores the
new setting.
ENTR
accepts the
new setting.
DIAG AIO
CONC_OUT_1 AUTO CAL: OFF
< SET
SET>
EDIT
EXIT
6.7.3.4. Analog Output Offset Adjustment
Some analog signal recorders require that the zero signal is significantly different from the
baseline of the recorder in order to record slightly negative readings from noise around the
zero point. This can be achieved in the M200E by defining a zero offset, a small voltage
(e.g., 10% of span), which can be added to the signal of individual output channels by
pressing SETUP-MORE-DIAG-ENTR-NEXT-NEXT and the following keys:
044100102 Rev A
65
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...