Model 200E Instruction Manual
Troubleshooting & Repair
Measure the voltage across TP1 and TP2, it should be 10.0
±
0.25 V. If not, the board is
faulty. Measure the voltage across the leads of capacitor C2. It should be 5.0 ± 0.25 V, if
not, the board may be faulty.
11.5.15.1. Reaction Cell Pressure
Measure the voltage across test points TP1 and TP5. With the sample pump disconnected or
turned off, the voltage should be 4500
±
250 mV. With the pump running, it should be 800-
1700 mV depending on the performance of the vacuum pump. The lower the reaction cell
pressure, the lower the resulting voltage is. If this voltage is significantly different, the
pressure transducer S1 or the board may be faulty. If this voltage is between 2 and 5 V, the
pump may not be performing well, check that the reaction cell pressure is less than
10 in-Hg-A (at sea level). Ensure that the tubing is connected to the upper port, which is
closer to the sensor’s contacts; the lower port does not measure pressure.
11.5.15.2. Sample Pressure
Measure the voltage across test points TP1 and TP4. With the sample pump disconnected or
turned off, this voltage should be 4500
±
250 mV. With the pump running, it should be
about 0.2 V less as the sample pressure drops by about 1 in-Hg-A below ambient pressure.
If this voltage is significantly different, the pressure transducer S2 or the board may be
faulty. A leak in the sample system to vacuum may also cause this voltage to be between
about 0.6 and 4.5. Make sure that the front panel reading of the sample pressure is at
about 1 in-Hg-A less than ambient pressure. Ensure that the tubing is connected to the
upper port, which is closer to the sensor’s contacts; the lower port does not measure
pressure.
044100102 Rev A
213
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...